BIEN AIROTOR - PRESTIGE L

Availability :
In Stock
Airotor handpiece
57275 57275
Quantity:


Check Availability at

Eligible for Cash on Delivery      Usually Delivered in 2-4 days

Easy Returns and Replacement

Payment Options (Credit Card, Debit Card, Net Banking, COD)

  • Type - Turbine with light, small head
  • Connection - Using 4-way Unifix quick-connect coupling, rotating 360 degree
  • Speed of rotation - 335,000 rpm
  • Recommended air pressure - 3.1 bar
  • Cooling - Diffuser, with 3 separate air and water sprays, water antiretraction valve
  • Chuck locking mechanism - Push-button with anti-heating system
  • Light transmission - Multi-strand optical glass conductor
  • Lighting - Halogen
  • Cleaning/Sterilisation - Thermo washer disinfectable / Steam autoclave up to 135 degree C
  • Weight - 73 g
  • Prestige L ref. - 1600380-001
  • Prestige L + 4-way Unifix set ref. - 1700184-001

 

MAINTENANCE OF TURBINES.

Preparation
A.

REMOVING THE INSTRUMENT

  • Remove the rotary instrument from the chuck mechanism
Tissue_rubbing
B.

PRE-DISINFECTION

  • External disinfection

Do not immerse the instrument in a disinfectant bath.

CLEANING
Cleaning_A
A.

CLEAN UNDER RUNNING WATER AT <38°C

Cleaning_B
B.

REMOVING DIRT/DEPOSITS

  • Carefully remove all traces of dirt or deposits from the spray nozzles using the Bien-Air cleaning wire
  • Dry the nozzles with compressed air then dry the rest of the instrument with a cloth

Do not immerse in an ultrasonic cleaner

Cleaning_CSpraynet-Tissue-spray_CTissue_rubbing_C
C.

INTERNAL AND EXTERNAL CLEANING WITH SPRAYNET

  • Shake the Spraynet before use
  • Use the Spraynet attachment 1
    (for 2-/3-way and 4-way turbines, see inset below)
  • Keep the tip of the instrument pointing down and protect with an absorbent cloth; press the Spraynet button down for 1 second
  • Clean the exterior of the turbine with a cloth soaked in Spraynet
FOR 2-/3-WAY AND 4-WAY TURBINES ONLY
Cleaning_C-primeBlack-Pearl-Eco_2-3waysBlack-Pearl-Eco_4-ways
C'.

CLEANING THE INSIDE WITH SPRAYNET FOR 2-/3-WAY AND 4-WAY TURBINES ONLY

  • Use the Spraynet attachment 2
  • Keep the tip of the instrument pointing down and protect with an absorbent cloth; press the Spraynet button down for 1 second

Spray into the opening as shown in the illustrations for “2-/3-way” and “4-way”

DISINFECTION
Disinfectant-Tissue-spray_ATissue_rubbing_A
A.

DISINFECTION USING SURFACE DISINFECTANT (follow the manufacturer's instructions for use)

Do not immerse the instrument in a disinfectant bath

OR
Thermal-disinfector
B.

THERMAL DISINFECTION (follow the manufacturer's instructions for use)

LUBRICATION
Lubrification_A1Lubrification_A2
A.

LUBRICATION WITH LUBRIMED

OR
Lubrification_B1Support_B
B.

LUBRICATION WITH LUBRIFLUID

  • Shake the Lubrifluid before use
  • Use the Lubrifluid attachment 23 ou 4depending on the type of instrument * (For 2-/3-way and 4-way turbines, see inset below)
  • Keep the tip of the instrument pointing down and protect with an absorbent cloth; press the Lubrifluid button down for 1 second
  • Then, run an operating test for 30 seconds at low speed to remove any surplus lubricant
  • Store immediately: store the instruments with the tip pointing downwards so that any oil residues can flow out
FOR 2-/3-WAY AND 4-WAY TURBINES ONLY
LubrificationBlack-Pearl-Eco_2-3waysBlack-Pearl-Eco_4-ways
B'.

INTERNAL LUBRICATION WITH LUBRIFLUID FOR 2-/3-WAY AND 4-WAY TURBINES ONLY

  • Use the Lubrifluid attachment 2
  • Keep the tip of the instrument pointing down and protect with an absorbent cloth; press the Lubrifluid button down for 1 second

Spray into the opening as shown in the illustrations for “2-/3-way” and “4-way”

STERILISATION
Sterilization
A.

PLACE THE TURBINE IN A STERILISATION BAG

Sterilizer
B.

STERILISE IN ACCORDANCE WITH THE MANUFACTURER'S INSTRUCTIONS

  • Instruments bearing the "135°C" symbol can be sterilised in an autoclave

Only sterilise instruments which have been cleaned and lubricated

 